ABC Article Directory banner displaying blue butterfly logo. Click to go directly to the main Homepage
Your Ad Here

Home | Technology | Gadgets And Gizmos

Add This Social Bookmark Button





animated blue butterfly symbol for the ABC Article Directory

Virginia's new breath alcohol tester: the Intox EC/IR II


By: Robert Keefer

On April 23, 2008, Virginia entered into a contract to purchase a new breath tester. On that same day, in accordance with its Regulations for Breath Alcohol Testing, Virginia began listing Intoximeter's EC/IR II with the Virginia test protocol as an approved breath test device.

Alka Lohmann is the chief of Virginia's unaccredited breath alcohol section. At the May 5, 2008 Scientific Advisory Committee meeting, Ms. Lohmann gave a status report on the EC/IR II. She reported that out of 50 states this tester is used in only four states besides Virginia: West Virginia, Wisconsin, North Carolina and Tennessee. Ms. Lohmann did not indicate why only 10% of the states used the EC/IR II. She stated that the new tester offered enhanced communication capabilities, increased trouble shooting form the lab and the ability to use a dry gas.

Ms. Lohmann was asked about the types of reports that can be requested related to the device. She stated that the same reports presently available will be available for the new machine.

One of the basic principles of analytical chemistry is the each analysis such be corroborated by an independent, alternative method. Agreement between the two different analytical methods insures the reliability of the results.

The name EC/IR II implies that the machine will measure the subject's alcohol concentration in two methods: EC (electrochemical or fuel cell) and IR (infrared technology). In so doing,the EC/IR II would insure reliability of its results by these different analytical methods. This internal corroboration would give the EC/IR II a distinct reliability advantage over devices utilizing only one technology.

Unfortunately, Intoximeter was unable to coordinate the different analytical methods. Because of that failure, the instrument does not use two independent methods to measure alcohol concentration. Instead, the EC/IR II uses only the fuel cell to measure alcohol and determine blood alcohol content. The IR is employed only to detect mouth alcohol through a slope detector.

The fuel cell component does not have the ability to detect mouth alcohol. The fuel cell merely measures total concentration of alcohol present whether from lung air or mouth.

Over time, fuel cells change in sensitivity and must be calibrated frequently to retain accuracy. IR detectors do not require calibration every few weeks. As the electrode microstructure changes over time the sensor baseline drifts. That drift takes the EC out of sych with the IR detector.

Calibrating the device every few weeks would be very demanding and time consuming. Some states have resolved that issue by disabling the IR detector entirely. Once the IR detector is disabled the device has no capacity for detecting mouth alcohol.

Copyright (c) 2008 Robert Keefer

Article Source: ABC Article Directory

www.duidriver.net www.bobkeeferlaw.com www.recklessdriving.net





Bookmark and Share

 eMail This Article to Friends

Please Rate this Article


Not yet Rated



Click the XML Icon Above to Receive Gadgets and Gizmos Articles Via RSS!



Copyright ABC Article Directory™ All rights protected. Script Services by: Sustainable Website Design
Use of our free service is protected by our Privacy Policy and Terms of Service  Contact Us
Creative Commons License
This work is licensed under a Creative Commons Attribution-No Derivative Works 3.0 Unported License.

Wind Powered Hosting

Powered by Article Dashboard